MTR8A Specifications PDF Print E-mail
The Manual Test Rig incorporates a variac connected to a power meter with either a single phase, or three phase test lead around the test core
Variac Input 
  • 380-440Vac phase-phase (3 phase), or 220-240Vac phase to neutral 50/60 Hz
  • Must have Neutral and Earth connection to mains supply for star test in three phase cores  
 Variac Output
  •  Auto transformer type, continuously variable output
  • 415Vac input (phase to phase) => 470Vac output (phase to phase) (19-12% step up in voltage)
  • 6.5kVA max power
  • 8 Amps per phase
  • All wiring rated to 10 Amps continuous duty cycle
 Power Meter

  •  Supply: 230Vac ± 10% 50/60Hz single phase 10VA
  • Voltage measurement: 0-500Vrms auto ranging
  • Current measurement: 0-15Arms direct with auto scaling
  • Power: 0-7.5kW auto ranging
  • Apparent Power (VA) 0-7.5kVA
  • Reactive Power (VAr) 0-7.5kVAr
  • Power Factor -1.0-> +1.0
 Test Cables
  •  rated to 500Vac
  •  10A pin contacts on plug / sockets
  • plug and play, no wiring by operator